FDA-Device2023-11-29Class II

Compress Device Short Anchor Plug, 20 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178562

Biomet, Inc.
Hazard

Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery

Sold states
Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, CANADA, CHILE, JAPAN, NETHERLANDS, NEW ZEALAND.
Affected count
N/A
Manufactured in
56 E Bell Dr, N/A, Warsaw, IN, United States
Products
Compress Device Short Anchor Plug, 20 MM-For Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. Item Number: 178562

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0383-2024