FDA-Device2018-01-24Class II
Phadia 1000 Instrument, Article Number 12-3800-01.
Phadia US Inc
Hazard
The "Retry" command does not function properly which could cause a shortage of Wash and Rinse solution and affect assay performance and test results.
Sold states
Nationwide
Affected count
70 units
Manufactured in
4169 Commercial Ave, N/A, Portage, MI, United States
Products
Phadia 1000 Instrument, Article Number 12-3800-01.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0387-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief