FDA-Device2012-11-28Class II

00789504001 VerSys Femoral Head Provisional 40mm -3.5 00789504002 VerSys Femoral Head Provisional 40mm +0 00789504003 VerSys Femoral Head Provisional 40mm +3.5 00789504004 VerSys Femoral Head Provisional 40mm +7 00789504005 VerSys Femoral Head Provisional 40mm +10.5 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty

Zimmer, Inc.
Hazard

Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat

Sold states
Worldwide distribution: USA (nationwide) and countries of: Asia, Barbados, Canada, Europe, Mexico, Puerto Rico, San Juan, and South America.
Affected count
128,946 all lots, all sizes
Manufactured in
345 E Main St, N/A, Warsaw, IN, United States
Products
00789504001 VerSys Femoral Head Provisional 40mm -3.5 00789504002 VerSys Femoral Head Provisional 40mm +0 00789504003 VerSys Femoral Head Provisional 40mm +3.5 00789504004 VerSys Femoral Head Provisional 40mm +7 00789504005 VerSys Femoral Head Provisional 40mm +10.5 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0394-2013