FDA-Device2018-01-31Class II

Brilliance iCT - Model 728306 Computed Tomography X-ray systems

Philips Medical Systems (Cleveland) Inc
Hazard

Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.

Sold states
Worldwide distribution, including US nationwide.
Affected count
559 total
Manufactured in
595 Miner Rd, N/A, Cleveland, OH, United States
Products
Brilliance iCT - Model 728306 Computed Tomography X-ray systems

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0403-2018

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief