FDA-Device2019-11-27Class II
Blanketrol III Hyper-Hypothermia System- body hypothermia system temperature management Model 233 - Product Usage: The system is composed of a heater, a compressor, a circulating pump and blankets/pads.
Cincinnati Sub-Zero Products LLC, a Gentherm Company
Hazard
Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modes
Sold states
Worldwide distribution - US Nationwide and countries of Afghanistan, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Belgium, Bhutan, Brazil, Brunei, Canada Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, France, Germany, Guatemala, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Nepal, Netherlands, New Zealand, Oman, Pakistan, Palestine, Panama, Philippines, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, Venezuela, Vietnam.
Affected count
3932
Manufactured in
12011 Mosteller Rd, N/A, Cincinnati, OH, United States
Products
Blanketrol III Hyper-Hypothermia System- body hypothermia system temperature management Model 233 - Product Usage: The system is composed of a heater, a compressor, a circulating pump and blankets/pads.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0417-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief