FDA-Device2022-12-14Class II

LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Used in spinal fixation surgery.

Integrity Implants Inc.
Hazard

Potential for intraoperative screw head separation at weld location.

Sold states
US distribution to California, Indiana, New York, Oklahoma, Washington DC
Affected count
50 screws
Manufactured in
354 Hiatt Drive, N/A, Palm Beach Gardens, FL, United States
Products
LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Used in spinal fixation surgery.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0421-2023