FDA-Device2012-11-28Class III

Minotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High). Minotrol 16 is a tri-level control intended for in vitro diagnostic use in monitoring the accuracy and precision on HORIBA Medical hematology blood cell counters using HORIBA Medical reagents. Minotrol 16 is an in vitro diagnostic reagent composed of human erythrocytes, mammalian leukocytes, and platelets in a plasma-like fluid with preservatives. Minotrol 16 is a stable material that prov

Horiba Instruments, Inc dba Horiba Medical
Hazard

HORIBA Medical Irvine Technical Support received reports from some customers that have experienced high recover values for Hemoglobin when using Minotrol 16 lot #MX090 (Level High). Customer communication was sent 10/01/2010.

Sold states
Nationwide Distribution
Affected count
22,954
Manufactured in
34 Bunsen, N/A, Irvine, CA, United States
Products
Minotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High). Minotrol 16 is a tri-level control intended for in vitro diagnostic use in monitoring the accuracy and precision on HORIBA Medical hematology blood cell counters using HORIBA Medical reagents. Minotrol 16 is an in vitro diagnostic reagent composed of human erythrocytes, mammalian leukocytes, and platelets in a plasma-like fluid with preservatives. Minotrol 16 is a stable material that provides a means of determining the accuracy and precision of hematology blood cell counters. It is handled in the same manner as a patient specimen. The assay tables are determined on validated instruments using the appropriate reagents

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0429-2013