FDA-Device2025-11-12Class IIPROCESSING DEFECT

Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer

OTHERNationwide distributionMedical devices
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
Hazard

An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.

Intended for
All ages
Illnesses reported
None reported
Sold states
Worldwide
Affected count
657 units
Manufactured in
1000 Lake Hazeltine Dr, N/A, Chaska, MN, United States
Products
Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0429-2026