MAGNETOM Cima.X Upgrade. Model Number: 11689304.

There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

From the FDA notice · Read the official notice ↗ · All Siemens recalls

What to do

The agency's notice has the remedy and who to contact.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
MAGNETOM Cima.X Upgrade. Model Number: 11689304.
Sold in
US distribution to California. International distribution to Australia, Germany, Great Britian.
Made for
All ages (0 months and older)

Details

Units affected
4 system (1 US; 3 OUS)
Illnesses reported
None reported
Manufactured in
40 Liberty Blvd, Malvern, PA, United States
Classification
Class I
Recall number
Z-0431-2026

In the FDA's words

Description
MAGNETOM Cima.X Upgrade. Model Number: 11689304. There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room. Model Number: 11689304. UDI: (01)04056869975245(21)237304. Serial Number: 237304. Distribution: US distribution to California. International distribution to Australia, Germany, Great Britian. Affected quantity: 4 system (1 US; 3 OUS)
Hazard
There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.