FDA-Device2015-12-23Class II

MiniMed 640G, 3.0mL mmol/L, 3.0mL mg/dL, Model No. MMT-1711 and MMT-1712 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous monitoring of glucose levels in the fluid under the skin, and possible low and high blood glucose episodes.

Medtronic MiniMed Inc.
Hazard

The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.

Sold states
Internationally Distribution only to the countries of : Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya, Luxembourg, Namibia, Netherlands, Norway, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom.
Affected count
1052 units
Manufactured in
18000 Devonshire St, N/A, Northridge, CA, United States
Products
MiniMed 640G, 3.0mL mmol/L, 3.0mL mg/dL, Model No. MMT-1711 and MMT-1712 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous monitoring of glucose levels in the fluid under the skin, and possible low and high blood glucose episodes.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0438-2016