FDA-Device2025-11-19Class IIPROCESSING DEFECT
Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
Hazard
The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.
Sold states
Worldwide distribution - US Nationwide and the countries of Canada, India, Taiwan, Germany, Australia, Singapore.
Affected count
1528 units
Manufactured in
3500 Corporate Pkwy, Center Valley, PA, United States
Products
Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones. Component: ShockPulse Lithotripsy System
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0452-2026More Olympus Corporation of the Americas recalls
- FDA-Device2026-04-08Olympus SOLTIVE Premium laser system recalled for defective power supply
- FDA-Device2026-04-08Olympus SOLTIVE Pro SuperPulsed Laser System recalled for power supply defect
- FDA-Device2026-04-01Olympus PKS Cutting Forceps recalled for defective welds causing jaw breakage
- FDA-Device2026-04-01Olympus MAJ-1443 and MAJ-1444 endoscope suction valves no longer compatible with OER-Pro and OER-Elite reprocessors
- FDA-Device2026-04-01Olympus PKS Cutting Forceps recalled for defective welds causing jaw breakage
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief