FDA-Device2025-11-19Class IIPACKAGING DEFECT

Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/26 to include: CARESCAPE B450 MBA313 2095800-001 CARESCAPE MONITOR B450 (ATO MODEL) 2068491-002 CARESCAPE MONITOR B450 B1 2107633-001 CARESCAPE MONITOR B450-GS (V2.0.8) 2094080-001

BATTERYNationwide distributionMedical devices
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
Hazard

Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide - US Nationwide distribution.
Affected count
3271
Manufactured in
Kuortaneenkatu 2, Helsinki, Finland
Products
Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/26 to include: CARESCAPE B450 MBA313 2095800-001 CARESCAPE MONITOR B450 (ATO MODEL) 2068491-002 CARESCAPE MONITOR B450 B1 2107633-001 CARESCAPE MONITOR B450-GS (V2.0.8) 2094080-001

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0453-2026