FDA-Device2015-12-30Class II

Alaris EtCO2 Module, Model 8300. A capnograph that continuously monitors end tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).

CareFusion 303, Inc.
Hazard

CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2. The effect to the EtCO2 module tested with a higher concentration of CO2 could lead to an incorrect measurement of CO2.

Sold states
US Distribution.
Affected count
157 units
Manufactured in
10020 Pacific Mesa Blvd, N/A, San Diego, CA, United States
Products
Alaris EtCO2 Module, Model 8300. A capnograph that continuously monitors end tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0455-2016