FDA-Device2015-12-30Class II
Reagent Management System (RMS) Dimension EXL with LM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿ system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex¿ reagent cartridges, allowing the Dimension¿ system to process samples without interruption.
Siemens Healthcare Diagnostics Inc.
Hazard
Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
Sold states
US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.
Affected count
106
Manufactured in
101 Silvermine Rd, N/A, Brookfield, CT, United States
Products
Reagent Management System (RMS) Dimension EXL with LM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿ system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex¿ reagent cartridges, allowing the Dimension¿ system to process samples without interruption.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0458-2016Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief