FDA-Device2015-12-30Class II

Reagent Management System (RMS) Dimension RxL Max w/o HM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿ system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex¿ reagent cartridges, allowing the Dimension¿ system to process samples without interruption.

Siemens Healthcare Diagnostics Inc.
Hazard

Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.

Sold states
US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.
Affected count
3
Manufactured in
101 Silvermine Rd, N/A, Brookfield, CT, United States
Products
Reagent Management System (RMS) Dimension RxL Max w/o HM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿ system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex¿ reagent cartridges, allowing the Dimension¿ system to process samples without interruption.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0460-2016