FDA-Device2018-11-21Class II
Arrow AGB+ Multi-Lumen CVC Kit, 7 Fr., 3 Lumen, 20cm catheter length, .025 inch dia. spring-wire guide, Product Code CDC-45703-B1A - Product Usage: The multiple-lumen catheter permits venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. The ARROWg+ard Blue Plus antimicrobial catheter is intended to help provide protection against catheter-related infections. It is not intended to be used as a treatment for existing infections nor is it indicated for
Arrow International Inc
Hazard
The products labeling contains inconsistencies. In certain instances, the lidstock, banner card and shipper label state that the Spring Wire Guide (SWG) within is .032 rather than the correct SWG size .025 . For other lots in scope the lidstock correctly states the SWG size .025 however the banner card states the incorrect SWG size .032 .
Sold states
US Nationwide distribution in the states of CA, FL, ID, IL, MI, NC, NJ, PA, and WI.
Affected count
3640
Manufactured in
2400 Bernville Rd, Reading, PA, United States
Products
Arrow AGB+ Multi-Lumen CVC Kit, 7 Fr., 3 Lumen, 20cm catheter length, .025 inch dia. spring-wire guide, Product Code CDC-45703-B1A - Product Usage: The multiple-lumen catheter permits venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. The ARROWg+ard Blue Plus antimicrobial catheter is intended to help provide protection against catheter-related infections. It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0468-2019Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief