NexGen Option Stemmed Tibial Component Size 6, Item Number 00-5986-047-02
Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0469-2023More Zimmer, Inc. recalls
- FDA-Device2026-07-01Zimmer Off-Axis Comprehensive Shoulder System reamer guide recalled for breakage risk
- FDA-Device2026-07-01Zimmer Off-Axis Alliance Glenoid Reamer Guide recalled for plastic breakage risk during surgery
- FDA-Device2026-07-01Zimmer Persona Revision Femoral Distal Augment recalled for wrong component in package
- FDA-Device2026-07-01Off-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444
- FDA-Device2026-07-01Off-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040242