FDA-Device2025-11-19Class II
Medline Renewal LigaSure Small Jaw Instrument (Purple/White), Compatible with FT10 Generator 18.8 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
Medline Industries, LP
Hazard
Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged.
Sold states
US Nationwide distribution.
Affected count
18,117 units
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Products
Medline Renewal LigaSure Small Jaw Instrument (Purple/White), Compatible with FT10 Generator 18.8 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed