FDA-Device2022-12-21Class II

BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267

Becton Dickinson & Co.
Hazard

Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for Haemophilus isolation in a clinical specimen.

Sold states
Worldwide distribution - US Nationwide and the countries of Canada, India, Philippines.
Affected count
9,600 EA
Manufactured in
7 Loveton Cir, N/A, Sparks, MD, United States
Products
BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0473-2023