FDA-Device2024-11-27Class II
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4704
Hobbs Medical, Inc.
Hazard
Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
Sold states
Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.
Affected count
6 units
Manufactured in
8 Spring St, N/A, Stafford Springs, CT, United States
Products
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4704
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0473-2025Don't want to check this manually?