FDA-Device2024-11-27Class II
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723
Hobbs Medical, Inc.
Hazard
Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
Sold states
Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.
Affected count
10 units
Manufactured in
8 Spring St, N/A, Stafford Springs, CT, United States
Products
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723
Official notice