FDA-Device2025-11-19Class IIPROCESSING DEFECT

GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

OTHERNationwide distributionMedical devices
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
Hazard

The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.

Intended for
All ages
Illnesses reported
None reported
Sold states
Domestic: US Nationwide Distribution. International: Australia, Belgium, Canada, Chile, China Mainland, Czech Republic, Denmark, Ecuador, Finland, France, Germany, GHANA, Hungary, India, Italy, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab Emir., Vietnam,
Affected count
236 units
Manufactured in
14 Electronics Ave, Danvers, MA, United States
Products
GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0480-2026