FDA-Device2015-12-30Class II

Artis systems with large display from Siemens Medical Solutions USA, Inc. Interventional Fluoroscopic X-ray for angiographic procedures.

Siemens Medical Solutions USA, Inc
Hazard

Due to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only sporadic cases have been observed in factory screening.

Sold states
US Distribution to the states of : IL, WI, MA, FL, AL, TN, ND, OH, TX, CA, NJ and MI.
Affected count
17
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Artis systems with large display from Siemens Medical Solutions USA, Inc. Interventional Fluoroscopic X-ray for angiographic procedures.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0481-2016