FDA-Device2025-11-19Class IIPROCESSING DEFECT
SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600
OTHERNationwide distribution
Hazard
Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.
Sold states
US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.
Affected count
56 units (55 US, 1 OUS)
Manufactured in
1001 Murry Ridge Ln, Murrysville, PA, United States
Products
SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0481-2026More Philips Respironics, Inc. recalls
- FDA-Device2026-04-08Philips Respironics Trilogy Evo O2 ventilator recalled for delayed obstruction alarm
- FDA-Device2026-04-08Philips Respironics Trilogy Evo Universal ventilator recalled for delayed obstruction alarm
- FDA-Device2026-04-08Philips Respironics Trilogy EV300 ventilator recalled for delayed obstruction alarm
- FDA-Device2026-04-08Philips Respironics Trilogy Evo ventilator recalled for delayed obstruction alarm
- FDA-Device2026-04-01Philips Respironics Trilogy Evo ventilator recalled for inaccurate breath volume delivery with certain nebulizers
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief