FDA-Device2018-11-28Class II

Ascom Telligence Nurse Call System

Ascom (US) Inc.
Hazard

There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.

Sold states
AL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX UT, VA, WI
Affected count
57 units
Manufactured in
9024 Town Center Parkway, N/A, Lakewood Ranch, FL, United States
Products
Ascom Telligence Nurse Call System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0486-2019

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief