FDA-Device2018-11-28Class II
Ascom Telligence Nurse Call System
Ascom (US) Inc.
Hazard
There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.
Sold states
AL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX UT, VA, WI
Affected count
57 units
Manufactured in
9024 Town Center Parkway, N/A, Lakewood Ranch, FL, United States
Products
Ascom Telligence Nurse Call System
Official notice