FDA-Device2018-02-14Class I
MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Philips North America, LLC
Hazard
Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
Sold states
No U.S. distribution, Internationally to: India, Ireland, Japan, Republic of Korea, Malaysia, Mexico, South Africa & United Kingdom
Affected count
71 units
Manufactured in
3000 Minuteman Rd, N/A, Andover, MA, United States
Products
MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Official notice