FDA-Device2018-02-07Class II

Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.

Tosoh Bioscience Inc
Hazard

This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.

Sold states
Worldwide Distribution - U.S. Nationwide - Latin America
Affected count
565 total devices U.S.
Manufactured in
3600 Gantz Rd, Grove City, OH, United States