FDA-Device2012-12-19Class II
Philips IntelliVue Information Center/Client and Philips IntelliVue Information Center iX The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its d
Philips Healthcare Inc.
Hazard
Internal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms.
Sold states
Worldwide Distribution - USA (nationwide) and Canada.
Affected count
3165
Manufactured in
3000 Minuteman Road, N/A, Andover, MA, United States
Products
Philips IntelliVue Information Center/Client and Philips IntelliVue Information Center iX The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0495-2013Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief