FDA-Device2024-11-27Class II

Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile

Abbott Medical
Hazard

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.

Sold states
New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.
Affected count
5 OUS
Manufactured in
177 E County Road B, N/A, Saint Paul, MN, United States
Products
Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0495-2025

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief