FDA-Device2019-11-27Class II

WECK AutoEndo5, Automatic Hem-O-Lok Clip Applier, REF AE05ML Automatic Endoscopic Surgical Ligating Clip Appliers are delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips

Teleflex Medical
Hazard

Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).

Sold states
Worldwide distribution, including US nationwide (including Puerto Rico).
Affected count
34491 devices
Manufactured in
3015 Carrington Mill Blvd, N/A, Morrisville, NC, United States
Products
WECK AutoEndo5, Automatic Hem-O-Lok Clip Applier, REF AE05ML Automatic Endoscopic Surgical Ligating Clip Appliers are delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0496-2020