Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS; 106524 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
From the FDA notice · Read the official notice ↗ · All Thoratec recalls
What to do
The agency's notice has the remedy and who to contact.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS; 106524 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the workload of the left ventricle, thereby restoring the patient's systemic perfusion while palliating the underlying pathology. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus all of the features, controls, attachments, interfaces, power sources, supporting equipment, labeling, and tools required to achieve the desired therapeutic benefit. Component: No
- Sold in
- Nationwide
- Made for
- Adults (18 years to 100 years)
This recall covers 4 products — see all of them.
Details
- Units affected
- 29059
- Illnesses reported
- None reported
- Manufactured in
- 6035 Stoneridge Dr, Pleasanton, CA, United States
- Classification
- Class II
- Recall number
- Z-0505-2026
In the FDA's words
- Description
- Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS; 106524 Software Version: N/A Product Description: The HeartMate 3" Left Ventricular Assist System (LVAS) is a set of equipment and materials that together comprise a medical device designed to provide therapeutic benefit to those affected with advanced heart failure. In service, the LVAS assumes some or all of the workload of the left ventricle, thereby restoring the patient's systemic perfusion while palliating the underlying pathology. The LVAS features a Left Ventricular Assist Device (LVAD), a blood pump intended for long-term implantation in such patients, an extracorporeal Controller, plus all of the features, controls, attachments, interfaces, power sources, supporting equipment, labeling, and tools required to achieve the desired therapeutic benefit. Component: No Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm. Lot Code: All units of the following models are impacted: Model 106524US UDI: 00813024013297 Model 106524INT UDI: 0081302401171 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used. Distribution: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan. Affected quantity: 29059
- Hazard
- Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
4 products in this recall
HeartMate LVAD controllers recalled for backup battery corrosion
Thoratec, a HeartMate company, is recalling four HeartMate heart pump system products because the backup battery cable connector may corrode. This can trigger a Backup Battery Fault alarm. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, He… (this page)Z-0505-2026 · 29059
- Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3…Z-0506-2026 · 35376
- Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, US…Z-0507-2026 · 248
- Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartM…Z-0508-2026 · 2424