FDA-Device2018-02-14Class II

ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.

Siemens Healthcare Diagnostics, Inc
Hazard

Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020).

Sold states
US and OUS.
Affected count
5992 units (US and OUS)
Manufactured in
333 Coney St, N/A, East Walpole, MA, United States
Products
ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0515-2018

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief