FDA-Device2025-11-26Class IIPROCESSING DEFECT

MEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202R; 2) ACDF CDS, REF CDS941061M; 3) NEURO PACK-LF, REF DYNJ0038440C; 4) NEURO VP SHUNT/VAGAL NERVE-RF, REF DYNJ21925S; 5) NEURO-ORTHO EXTRAS, REF DYNJ44882M; 6) LAMINECTOMY, REF DYNJ903870F.

ELECTRICAL SHOCKFIRENationwide distributionMedical devices
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
Hazard

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

Intended for
Adults
Illnesses reported
None reported
Sold states
US Nationwide distribution.
Affected count
1315 kits
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Products
MEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202R; 2) ACDF CDS, REF CDS941061M; 3) NEURO PACK-LF, REF DYNJ0038440C; 4) NEURO VP SHUNT/VAGAL NERVE-RF, REF DYNJ21925S; 5) NEURO-ORTHO EXTRAS, REF DYNJ44882M; 6) LAMINECTOMY, REF DYNJ903870F.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0522-2026

More Medline recalls

Want us to check the next one for you?

We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Free · One email a week · Unsubscribe anytime