FDA-Device2014-12-17Class II
Medtronic MiniMed Silhouette and/or MiniMed Sure-T Infusion Sets; Paradigm Silhouette infusion sets: MMT-368, MMT-368600, MMT-371, MMT-373, MMT-377, MMT-377600, MMT378, MMT-378600, MMT-38 1, MMT-381 600, MMT-382, MMT-382600, MMT-383, MMT-383600, MMT384, & MMT384600. Sure-T infusion sets MMT-862, MMT-863, MMT-864, MMT-865, MMT-866, MMT-873, MMT-874, MMTMMT-875, MMT-876, MMT-884, & MMT-886. MiniMed Duo sensor/infusion sets: MMT-5001A23, MMT-5001A43, MMT-5001B23, MMT-5001B43, MMT-5011A23, MMT-5011A
Medtronic MiniMed Inc.
Hazard
Tubing may detach at the connect/disconnect location on the MiniMed Silhouette and MiniMed Sure-T infusion sets. Detachment causes interrupted insulin delivery and the pump will not alarm to notify your patient. This can lead to hyperglycemia, and diabetic ketoacidosis (DKA).
Sold states
Worldwide distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lativa, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Venezuela.
Affected count
US 12,400,098 and Internationally 8,266,733
Manufactured in
18000 Devonshire St, N/A, Northridge, CA, United States
Products
Medtronic MiniMed Silhouette and/or MiniMed Sure-T Infusion Sets; Paradigm Silhouette infusion sets: MMT-368, MMT-368600, MMT-371, MMT-373, MMT-377, MMT-377600, MMT378, MMT-378600, MMT-38 1, MMT-381 600, MMT-382, MMT-382600, MMT-383, MMT-383600, MMT384, & MMT384600. Sure-T infusion sets MMT-862, MMT-863, MMT-864, MMT-865, MMT-866, MMT-873, MMT-874, MMTMMT-875, MMT-876, MMT-884, & MMT-886. MiniMed Duo sensor/infusion sets: MMT-5001A23, MMT-5001A43, MMT-5001B23, MMT-5001B43, MMT-5011A23, MMT-5011A43, MMT-5011B23, & MMT-5011B43. Silhouette and Sure-T disposable infusion sets are used with Medtronic insulin infusion pumps using Paradigm P-cap connectors. The infusion set connects the insulin reservoir in the pump to the patients body.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0524-2015Don't want to check this manually?