FDA-Device2018-02-14Class II

Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601, Model # (UPN) H7879006010, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

Angiodynamics, Inc.
Hazard

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Sold states
Worldwide distribution - US Nationwide
Affected count
14,353 units in total
Manufactured in
603 Queensbury Ave, N/A, Queensbury, NY, United States
Products
Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601, Model # (UPN) H7879006010, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0527-2018