FDA-Device2013-12-25Class II

Medfusion¿ 4000 Syringe Infusion Pumps Medfusion¿ 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes.

Smiths Medical ASD, Inc.
Hazard

Smiths Medical is conducting a recall involving all Medfusion¿ 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC po

Sold states
Worldwide Distribution - USA (nationwide ) including the states of : AZ, NY, TN, MS, AR, WV, VA, VT, PA, DE, DC, OH, CO, NH, MI, CA, FL, CT, WA, MD, NM, RI, MA, TX, NC, IL, KY, WI and MI., and the countries of CANADA, SAUDI ARABIA and AUSTRALIA.
Affected count
9560 (8601 US, 959 OUS)
Manufactured in
1265 Grey Fox Rd, N/A, Saint Paul, MN, United States
Products
Medfusion¿ 4000 Syringe Infusion Pumps Medfusion¿ 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0540-2014