FDA-Device2022-02-09Class II

ACUSON Sequoia Ultrasound Imaging System with Software Version VA25A/B

Siemens Medical Solutions USA, Inc.
Hazard

Due to a calculation error in the measurement when using 2D trace (manual trace) tool. The trace circumference value is overestimated and may potentially result in misdiagnosis of a patient's condition.

Sold states
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CN, DE, FL, GA, HI, IL, IN, IA, KS, KY, LA, MD, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY. The countries of United Arab Emirates, Armenia, Angola, Austria, Australia, Azerbaijan, Bangladesh, Belgium, Bahrain, Brazil, Switzerland, Chile, China, Colombia, Cyprus, Czech Republic, Germany, Denmark, Dominican Rep., Egypt, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Indonesia, Israel, Iceland, Italy, Japan, Republic Korea, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Poland, Portugal, Qatar, Romania, Serbia, Russian Fed., Saudi Arabia, Sweden, Singapore, Slovakia, Thailand, Turkey, Taiwan, Ukraine, South Africa.
Affected count
1161 systems
Manufactured in
22010 SE 51st St, N/A, Issaquah, WA, United States
Products
ACUSON Sequoia Ultrasound Imaging System with Software Version VA25A/B

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0547-2022