FDA-Device2018-12-12Class II
Captus 700 Thyroid Uptake System, Model # 5430-3137 and 5430-3138
Capintec Inc
Hazard
There is a potential for the spring arm failure as a result of a broken tension rod (a component inside the arm). A break in the tension rod can casue the collimator to fall downward to its lowest point of travel, approximately 25 inches from the ground.
Sold states
Worldwide distribution - US nationwide distributed in the states of AR, AZ, CA, CO, CT, DC, FL, IA, IN, LA, MD, MI, MN, NC, NJ, OH, OK, PA, SC, SD, TX, VA, WI, WV, and countries of Argentina, Brazil, Canada, China, Cyprus, Greece, Hong Kong, India, Japan, South Korea, Vietnam.
Affected count
65
Manufactured in
7 Vreeland Road, N/A, Florham Park, NJ, United States
Products
Captus 700 Thyroid Uptake System, Model # 5430-3137 and 5430-3138
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0551-2019Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief