FDA-Device2019-01-09Class I

StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instructions for cranial procedures using optical localization StealthStation Cranial version 3.0, 3.1 CE 0344 Rx Only 2018-02 9735489 Revision 3 Medtronic Navigation, Inc. Biopsy Needle Kit (9733068) Note: This device is an accessory to your StealthStation System. . . The Biopsy Needle is a presterilized , single-use, side-cutting needle for cranial tissue biopsies. CE 0344 Rx Only 20

Medtronic Navigation, Inc.
Hazard

An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.

Sold states
US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY OUS: Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Moldova, Morocco, Myanmar, Nepal, Netherland, New Zealand, Nicaragua, Norway, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam
Affected count
5487 impacted systems
Manufactured in
826 Coal Creek Cir, N/A, Louisville, CO, United States
Products
StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instructions for cranial procedures using optical localization StealthStation Cranial version 3.0, 3.1 CE 0344 Rx Only 2018-02 9735489 Revision 3 Medtronic Navigation, Inc. Biopsy Needle Kit (9733068) Note: This device is an accessory to your StealthStation System. . . The Biopsy Needle is a presterilized , single-use, side-cutting needle for cranial tissue biopsies. CE 0344 Rx Only 2017-05 9733068-01 Medtronic Navigation, Inc.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0552-2019

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.