FDA-Device2018-12-12Class II
Biomet 20mm Compress Device Short Anchor Plug With Drill Set / With Pin Inserter Ion Implant Item number: 178562 Product Usage: 1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. 2. Tumor resections. 3. Revision of previously failed total joint arthroplasty. 4. Trauma. The Compress¿ Segmental Femoral Replacement System (Short Spindle and Anchor Plug) components are intended for uncemented use only.
Zimmer Biomet, Inc.
Hazard
Potential metal burrs in the holes of the device preventing the appropriate drill and transverse pins from being inserted
Sold states
US Nationwide Distribution in the states of IN, DC, MD
Affected count
6 units
Manufactured in
56 E Bell Dr, N/A, Warsaw, IN, United States
Products
Biomet 20mm Compress Device Short Anchor Plug With Drill Set / With Pin Inserter Ion Implant Item number: 178562 Product Usage: 1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. 2. Tumor resections. 3. Revision of previously failed total joint arthroplasty. 4. Trauma. The Compress¿ Segmental Femoral Replacement System (Short Spindle and Anchor Plug) components are intended for uncemented use only.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0553-2019Don't want to check this manually?