FDA-Device2024-12-04Class II
Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
Shimadzu Corp. Analytical & Measuring Instruments Div.
Hazard
There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.
Sold states
Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.
Affected count
179
Manufactured in
1 Chome 1, Nishinokyokuwabara-Cho; Nakagyo; Hon-Machi; Higashiyama, Kyoto, Japan
Products
Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0553-2025Don't want to check this manually?