FDA-Device2018-02-14Class II

TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00

Mindray DS USA, Inc. dba Mindray North America
Hazard

A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.

Sold states
Distributed in 26 states: AL, CA, CO, FL, GA, IA, IN, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, WA, and WY; and Puerto Rico.
Affected count
203
Manufactured in
800 Macarthur Blvd, N/A, Mahwah, NJ, United States
Products
TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00