FDA-Device2012-12-26Class II
Persona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size H Left; Size H Right This device is used with the Persona Knee System for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o
Zimmer, Inc.
Hazard
All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening
Sold states
Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.
Affected count
189
Manufactured in
345 E Main St, Warsaw, IN, United States
Products
Persona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size H Left; Size H Right This device is used with the Persona Knee System for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0555-2013Don't want to check this manually?