FDA-Device2020-12-16Class II
Trinias, Digital Angiography System containing a MH-200, Ceiling Suspended C-arm Support - Product Usage: intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures.
Shimadzu Medical Systems
Hazard
The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement lever. This could result in possible collision or harm to the patient or system operator.
Sold states
US Nationwide distribution including in the states of KY, CA, MI, OH, CT, MI, NY, SC, NC,IN, IL, TN, LA, CT, MT.
Affected count
23 units
Manufactured in
20101 S Vermont Ave, N/A, Torrance, CA, United States
Products
Trinias, Digital Angiography System containing a MH-200, Ceiling Suspended C-arm Support - Product Usage: intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0555-2021Don't want to check this manually?