FDA-Device2020-12-16Class II

MANI Infusion Cannula 23G S - Product Usage: for use in injecting perfusate or aspirating out intraocular materials during ophthalmic surgery Ref: MTC 23S

Mani, Inc. - Kiyohara Facility
Hazard

Dimensions of the Luer connector parts out of specification, may cause leakage of fluid and unintended disconnection during use

Sold states
US Nationwide distribution in the state of MO.
Affected count
60 units (US)
Manufactured in
Park 8-3, Kiyoharakogyodanchi; Kiyoharakogyodanchi; Kiyoharakogyodanchi, Utsunomiya, N/A, Japan
Products
MANI Infusion Cannula 23G S - Product Usage: for use in injecting perfusate or aspirating out intraocular materials during ophthalmic surgery Ref: MTC 23S

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0558-2021