FDA-Device2023-12-27Class II
Philips Azurion systems with monoplane fixed ceiling mounts: 722078 Azurion 7 M12 722079 Azurion 7 M20 722223 Azurion 7 M12 722224 Azurion 7 M20 722227 Azurion 5 M12 722228 Azurion 5 M20
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Hazard
Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
Sold states
Worldwide - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, CANADA, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Latvia, Latvia, Republic of Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen.
Affected count
3821 units
Manufactured in
Veenpluis 4-6, N/A, Best, N/A, Netherlands
Products
Philips Azurion systems with monoplane fixed ceiling mounts: 722078 Azurion 7 M12 722079 Azurion 7 M20 722223 Azurion 7 M12 722224 Azurion 7 M20 722227 Azurion 5 M12 722228 Azurion 5 M20
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0559-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief