FDA-Device2020-12-16Class II
Evolut (TM) PRO+ System. With the following model numbers: a. EVPROPLUS-23US, Labeled as VLV EVPROPLUS-23US COMM JPN, VLV EVPROPLUS-23 COMM US; b. EVPROPLUS-26US, Labeled as VLV EVPROPLUS-26US COMM JPN, VLV EVPROPLUS-26 COMM US; c. EVPROPLUS-29US, Labeled as VLV EVPROPLUS-29US COMM JPN, VLV EVPROPLUS-29 COMM US; d. EVPROPLUS-34US, Labeled as VLV EVPROPLUS-34 COMM US, VLV EVPROPLUS-34 COMM US SA.
Medtronic Heart Valves Division
Hazard
Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).
Sold states
Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Russian Federation, Hungary, Kazakhstan, Luxembourg, Portugal, Norway, Egypt, Saudi Arabia.
Affected count
19,961 units
Manufactured in
1851 E Deere Ave, N/A, Santa Ana, CA, United States
Products
Evolut (TM) PRO+ System. With the following model numbers: a. EVPROPLUS-23US, Labeled as VLV EVPROPLUS-23US COMM JPN, VLV EVPROPLUS-23 COMM US; b. EVPROPLUS-26US, Labeled as VLV EVPROPLUS-26US COMM JPN, VLV EVPROPLUS-26 COMM US; c. EVPROPLUS-29US, Labeled as VLV EVPROPLUS-29US COMM JPN, VLV EVPROPLUS-29 COMM US; d. EVPROPLUS-34US, Labeled as VLV EVPROPLUS-34 COMM US, VLV EVPROPLUS-34 COMM US SA.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0561-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.