FDA-Device2018-02-21Class II

Vado Bi-Directional Steerable Sheath 8.8F, REF SS8FMB74, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.

Kalila Medical
Hazard

Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process.

Sold states
Distribution was made to DC, FL, GA, MD, MI, NJ, NY, OH, PA, SC, and VA. There was also military/government distribution. There was no foreign distribution.
Affected count
261 sheaths
Manufactured in
745 Camden Ave Ste A, Campbell, CA, United States
Products
Vado Bi-Directional Steerable Sheath 8.8F, REF SS8FMB74, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0564-2018