FDA-Device2014-01-01Class II
Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The HeartStart MRx EtCO2 Option monitors carbon dioxide (CO2) and measures end-t
Philips Medical Systems, Inc.
Hazard
MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg
Sold states
US Nationwide and OUS.
Affected count
18,529 units
Manufactured in
3000 Minuteman Rd, N/A, Andover, MA, United States
Products
Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The HeartStart MRx EtCO2 Option monitors carbon dioxide (CO2) and measures end-tidal carbon dioxide (EtCO2). The option provides an EtCO2 value and a CO2 waveform on the devices display.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0570-2014Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief