FDA-Device2014-01-01Class II
Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
Data Innovations, Inc.
Hazard
When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error
Sold states
Nationwide Canada and Britain
Affected count
24 systems
Manufactured in
120 Kimball Ave Ste 100, N/A, South Burlington, VT, United States
Products
Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
Official notice